Improvements in The Grating Drive of the Main...(1Mar94) Spectrometer (from NSO, NOAO Newsletter No. 37, 1 March 1994) There are two gratings mounted back to back in the main spectrometer: a 16.5 x 12.5 inch replica grating used to cover the spectrum 300 nm to 2.5 um and the 18.5 x 14.5 inch original grating which extends the spectrum to 12.5 um. The spectrum is scanned with a screw drive pulling on a steel tape wrapped around the spool containing the gratings. Photographs of the hyperfine structure of Hg 546.1 nm demonstrate a resolving power of about one million. However, when used in the scanning mode, as we normally use the spectrometer, it is much less - the linearity and smoothness of the drive are slightly worse than we had hoped and need - errors of 1 x 10**-4 to 5 x 10**-4 nm. To reduce the error, we have redesigned the tape-drive and are installing a new high-precision lapped screw and nut in February 1994. Keith Pierce