Next: 2.4 The 2.1-m GoldCam Spectrometer
Previous: 2.2 4-m RC Spectrograph
The ``CryoCam" is a
high-throughput analog of the RC Spectrograph. The instrument
actually consists of the RC Spectrograph but with the grating replaced
by any of 6 ``grisms" as the dispersive elements and with the
collimator mirror replaced by a collimating lens. The detector is a
dedicated Loral (Ford) 800 X 1200 device with relatively good
cosmetics sitting
in a fast (f/1) camera-dewar combination. The result of these all-transmission
optics is a total system throughput (telescope + spectrograph + CCD) that
is typically 20% (29% peak under excellent seeing conditions). Glass
optics curtail the throughput below 4000Å.
CryoCam is often used with
the multi-slit mechanism (see below); in this mode the acquisition is
done by interrupting the light path with a pick-off mirror below the
mask assembly. The entire 5
field can be viewed with this camera
when
the masks are in their out position.
Of course, during the actual exposure
the TV pickoff mirror must be out of the beam and so one is observing
``blind", but with
the excellent guiding provided by the auto-guider this is, in practice, no
problem. The instrument rotator at the 4-m allows remote control of the position angle of the slit on the sky, although such rotations need to be
done with the telescope within 30
of zenith.
Thus the spectrograph can be
rotated to the
parallactic
angle (to keep all of the light down the slit), or, if you need to
use a particular position angle, you can use the atmospheric dispersion
corrector (``Risley prisms"); see Sec. 4.4.1.
The table below shows the six grisms, the resulting wavelength coverage, dispersion, and resolutions. A table of the various offset slits can be found in Sec. A.1.2.
Notes: (a) The central wavelength changes
by up to 7% depending upon location in the field (perpendicular to the
slit). (b) Resolution
is based upon a FWHM of 3.5 pixels, corresponding to a 2.5-3.0
The RC and CryoCam spectrographs take 3.5-in square filters used for
order separation (see Sec. B). The following table gives
the available filters; transmission curves can be found in
Fig. 4.
Notes: (a) CuSO

slit. (c) Grism 810 covers an entire
octave; choice of wavelength coverage from 4300->10000Å is dictated by
blocking filter (i.e., a BG-39 will provide <4000-6000Å coverage without
overlapping; using an OG530 filter would allow coverage from approximately
5500Å to >10000Å. Note, however, that Grism 730 would be more
appropriate
in the latter case, providing the useful wavelength coverage but at
higher spectral resolution. (d) Glass optics curtail the throughput below
4000Å.
2.3.1 Blocking Filters for the 4-m RCSpec and CryoCam
Blocking Filters for RCSpec & CryoCam
Long
cut-off
Short cut-off
Filter
Thickness
Filter
Thickness
Filter
Thickness CuSO4a
8mm
WG-345
1mm
OG-515b
3mm BG-38
2mm
WG-360
2mm
OG-530
3mm BG-39
2mm
GG-375
1mm
OG-550b
3mm 4-96
5mm
GG-385
3mm
OG-570
3mm KG-2
2mm
GG-400b
3mm
OG-590b
3mm KG-3
2mm
GG-420
3mm
BG-610
3mm GG-455
3mm
GG-645
3mm GG-475
3mm
RG-695
3mm GG-495
3mm
RG-830
3mm
is too narrow for multi-slit work, but does cover
the entire length of the slit.
(b) New filter on-order; should be available
in early 1997.
Figure 4: The transmission curves for the various
RC and CryoCam blocking filters. You may examine these electronically
at http://www.noao.edu/kpno/filters/filters.html.
Next: 2.4 The 2.1-m GoldCam Spectrometer
Previous: 2.2 4-m RC Spectrograph
Updated: 02Sep1996