Next: A.2 Focusing the Spectrograph
Previous: A Setting Up and Focusing the Spectrographs
For the RC Spectrograph, the user can specify the central wavelength desired directly in Angstroms using the 4-m Telescope Control System (TCS) Graphical User Interface (GUI). The user must specify the grating name, the order, and the desired central wavelenth.
We give in the table below the encoder units corresponding to a particular wavelength setting. This table may be useful for simply substantiating the setup, and in determining the anamorphic demagnification, as explained in Sec. 4.3.
The wavelength in the center of the chip is determined by the choice
of grism. However, there are a number of slit inserts that
one can choose that will in fact change the central wavelength by
The grating tilt of the GoldCam spectrometer must be set by hand; care must
be taken not to overtighten the tilt clamp and observers are cautioned to
call for help if an adjustment is needed.
The table below shows the encoder setting for a particular wavelength.
(The grating tilt of
the zeroth order is 25.93.)
In reading the dial setting for the grating tilt, it is the number that
is covered by the pointer that gives the integral number of degrees.
A.1.2 CryoCam
of the wavelength coverage. The table below shows what offset slit aperture
plates are available, and what these shifts would be with the 300 l/mm
grisms. Other grisms will scale proportionally.
CryoCam Offset Slits Offset
Shift for 300 l/mm
Slit widths available 15.0mm
blue
-520Å
1.4'', 1.7'', 2.5'', 3.2'', 10.2''
10.8'', 11.1'', 21.0'' 7.5mm
blue
-260Å:
1.7'', 2.5'', 3.2'' 0.0mm
---
0Å
1.0'', 1.7'', 2.5'', 3.2'', 3.7'' 7.5mm
red
+260Å
1.7'', 2.5'', 3.2'' 11.0mm
red
+380Å
2.5'', 3.2'' 15.0mm
red
+520Å
1.7'', 2.5'', 3.2'' 22.6mm
red
+700Å
2.5''
A.1.3 GoldCam

Next: A.2 Focusing the Spectrograph
Previous: A Setting Up and Focusing the Spectrographs
Updated: 02Sep1996