Next: The Gold Spectrograph
Previous: System Overview
Figure 2 shows the measured throughput for the system using the
old Texas Instrument chip. The throughput with the Ford chip should be
significantly better at all wavelengths, (
20-50%).
The ordinate gives the fraction of photons striking the primary
mirror that are detected by the CCD. These measurements were made with the old
TI chip
by observing Gunn & Oke standards through a very wide slit (
11").
The falloff in the UV is mostly due to transmission losses in
the camera, while the loss of response towards 1 micron reflects the
decreasing sensitivity of the TI CCD
and the difference in blaze of the gratings used.
A practical working wavelength limit is
9700Å.
Figure 3 shows the count rate expected at the zenith for a 10th magnitude star observed with several gratings using the old TI5 chip.
Again, these refer to measurements taken with a very wide slit (
11").
Observed flux through a narrow slit will
depend critically on seeing.
The expected count rate can be computed from the above curves and the
number of photons per Angstrom per second incident on the 2.1-meter primary
mirror. The equation below gives the number of photons/sec/Å,
for a star of magnitude
and an extinction
loss of
magnitudes for the 2.1-meter telescope.
in
Å.

A star of
= 15.0 will illuminate the primary with
37
photons/second/Å at 4500 Å at the zenith (extinction at
4500 Å
0.2). For more discussion
see the end of the paper by Massey, Strobel, Barnes, and Anderson
Ap.J.
328
,315,1988.
Finally, differential
atmospheric refraction is non-negligible when observing in the
UV with a narrow slit.
For example, when working at an airmass of 2.0, there can be a displacement
of 2 arcseconds between images at 4000 Å and 8500 Å.
The spectrograph slit
can be rotated to the parallactic angle to reduce refractive effects, but
there is non-negligible overhead in spectrograph rotation (done
manually).