Next: Focusing Higher Dispersion Gratings
Previous: Set-up Checklist
Focusing the spectrograph is done by moving the collimator.
Auto-collimation is at 500, and offsets from the nominal focus
up to
100 units have been
noticed for different gratings. Steps of 20 units starting at higher
numbers are recommended for focusing.
By moving the collimator, however, one will be operating
slightly out of the auto-collimation position. The grating
will then be illuminated in a slightly converging or diverging beam, rather
than by parallel light. In this mode,
the resolution will be degraded slightly, but the
effect is probably not noticeable in this instrument where resolution is low
to start with.
(If focus offsets become large, say several hundred units, then
refocusing the CCD behind the camera will be necessary.)
With a 50 - 80
wide slit, one should be able to achieve a FWHM of
pixels when in focus.
When using the higher dispersion gratings, say
831 l/mm, the line
profiles may exhibit some asymmetry with FWHM being slightly greater
than expected. This is normal for gratings used at high inclinations.
See Section 5.5 for information on using gratings at higher dispersions.
The following steps can be used to focus:
, and check the spectral focus.
The telescope can be focused in this way also. Use the continuum of the star in place of the apertures. Try to duplicate the FWHM of the quartz.