System Design Notes
1. Introduction
This document describes the light source to be used in testing GNIRS on the Flexure Test Rig.
2. Overview
The flexure test
rig under construction is to evaluate the performance of GNIRS through the full
range of gravity vector orientations that it is likely to experience in actual
operation. While this testing will
cover many facets (mechanism operation, electronics cooling, thermal stability,
etc.), it will also be necessary to evaluate the optical performance of the
instrument at different orientations to demonstrate compliance with the flexure
requirements.
Three types of
flexure can be evaluated:
3.
Requirements
4. Possible
Solutions
4.1 Light Projector
We have in hand a
light projector used for the alignment of SQIID. It is a small integrating sphere into which a small
incandescent bulb is installed. This
gives a spectral continuum whose intensity can be easily controlled.
The resultant change in color with intensity is not a problem.
The integrating sphere has an output aperture of 20 mm and provision for
installing shim stock slides (33 x 50 mm x .010”) in which the desired target
have been cut. These may be either
fabricated in-house or by the vendor used by the Observatory for cutting slit
masks.
4.2 Translation
Stage
The projector used
for the SQIID alignment is mounted on a stage capable of X-Y motion using a
rack/pinion microscope stage and focus over 25 mm with a micrometer stage.
This arrangement is compact and lightweight, and seems to be sufficiently
robust to permit full motion with respect to the gravity vector without
exceeding the flexure requirements.
4.3
Optics
A refractive design
has advantages in simplicity, since it would permit a tube structure with a
mounting flange and perhaps gussets for structural stiffness.
Depending on the tolerances from the optical design, the lens mount could
be fairly simple, since it will not be required to withstand cryogenic
operation. The lenses would not
require anti-reflection coatings, although it may be desirable for high-index
materials.
We have designed a
triplet refractive solution that appears to meet the above requirements,
utilizing ZnSe, BaF2, and CaF2 lenses.
This provides 1:1 imaging of the source onto the input focal plane of the
instrument and has a total length of ~1200 mm, so it should project no more than
~300 mm above the beam of the flexure rig.
A 20 mm diameter stop 300 mm from the source will be re-imaged by the
lenses to the apparent distance of the Gemini secondary (16 m) as seen from the
focal plane and provide f/15 input to the instrument.
The geometric spot diagram yields essentially complete containment within
a 40-micron diameter aperture over the 1.0 – 2.4 micron wavelength range and
the 20 mm diameter field of the source. Over
the 3.0 – 4.2 microns range, the spot diameter is less than 60 microns,
smaller than the Airy diameter.
Maximum clear
aperture for the lenses is 63 mm. Allowing
the normal 10% overage for mounting, etc., and specifying that all three lenses
have the same diameter for ease of mounting will give a physical diameter 75 mm
or smaller, so the cost of the lenses should not be excessive.
Since one of the lenses (ZnSe) has a high index, we will probably specify
a quarter-wave AR coating centered at 1.6 microns.
5.
Test Plan
5.1 Instrument Focus
While there are
other ways to carry out this measurement, one can determine the position of the
instrument focus with respect to the ISS mounting plate by adjusting the focus
of the light source to obtain the best focus on the GNIRS slit, removing the
instrument, and measuring the position of the projected image.
One would presumably carry out this experiment at the very end of a
testing sequence, when the instrument would be coming off the flex rig in any
case. If the light source utilizes
refractive optics, it will be necessary to apply a small correction to the
observed optical focus location.
5.2 Instrument
Motion
This measurement
refers to the bulk motion of the instrument focal plane with respect to the ISS
as a function of gravitational loading. Since
some motion (on the order of 1 mm) is anticipated and unavoidable, this
experiment would simply provide a measure of this effect, although an unusually
large or discontinuous deflection would be cause for concern.
The experiment would entail centering the projected spot in a small
aperture with the instrument at zenith, then measuring the motion of the light
source necessary to recenter the spot with the instrument at different
orientations. This could be done
with the acquisition mirror in place. Any
motion of the image of the aperture on the detector would be a result of flexure
in the collimator, acquisition mirror, or camera (see next section).
It would be necessary to correct for the flexure of the projector by
measurements of the projected image with the instrument off the flex rig, which
can be carried out as part of the focus check (5.1).
5.2.1
Pupil Motion
With the instrument
pointing at the zenith, one may adjust the support truss to center the pupil
image of the projector onto the internal cold stop, while in pupil viewing mode.
Any motion of the cold stop with respect to the pupil image can then be
measured as a function of gravity load, after any effects of projector flexure
have been removed.
5.3 Internal Flexure
This important
component deals with any motion in the optical axis between the slit and the
detector, that is, one that would cause a shift of the re-imaged slit on the
array greater than the requirement of 0.1 pixel for a 15-degree motion of the
instrument tilt. One does not
require a focused point image on the slit for these measurements, and the use of
a large projected spot that overfills the slit will permit evaluation of this
effect without having to deal with the bulk motion of the instrument.
Use of an emission line source for these tests will yield the flexure in
both the spectral and spatial directions. All
scientifically viable combinations of grating and cross-dispersers should be
evaluated. The thermal continuum from the emission lamps overwhelms the
emission line spectrum beyond 3 microns, so investigation of flexure using the
red cameras may require working at shorter wavelengths and readjusting the
detector focus. This should not be
a problem, since all of these measurements are differential in nature.
Alternatively, it may be possible to utilize strong telluric absorption
lines in the 1 m atmospheric path between the light source and the GNIRS window.
5.4
Differential Flexure
Differential
flexure between the GNIRS slit and the OIWFS detector will result in a false
command to recenter the telescope pointing.
This flexure may be measured by using a target plate with two spots in
the projector. One of these,
representing the object, would be centered on the slit; the other would be
centered in the OIWFS target box using the gimbal mirror.
Any differential motion can be measured directly as a function of the
instrument orientation, after readjustment of the projector to recenter the
target on the slit after a change in orientation.
6.
Other Tests
This projector will
be useful for other characterization tests unrelated to flexure, such as setting
up the user grating positions, evaluation of spectral resolution and scattered
light, tests of the polarimetry modes (using a linear polarizer in front of the
projector), etc. We may wish to
carry out all instrument testing in the flexure rig bay, since all of the
infrastructure (cryocooler compressors, workstations, etc.) will be in place in
that location. In such a case, it
may be desirable to design a fixture for attaching the light projector to the
instrument truss so that optical tests in a fixed (presumably horizontal)
orientation can be carried out without the need to install the instrument on the
flexure rig.
R. Joyce
J. Elias
18 January 2002
Home | Management | System Design Notes | Requirements | Configuration Images | Subsystems | Archives | Staff | FTP Site | Internal Information
If
you have any questions or suggestions regarding this website, please contact Melissa Bowersock.
National Optical Astronomy Observatories,
950 North Cherry Avenue, P.O. Box 26732, Tucson, Arizona 85726,
Phone: (520) 318-8000, Fax: (520) 318-8360