SDN013.02 – Coefficient of Thermal Expansion

Revision C

1. Introduction

For the design of GNIRS, certain information on thermal expansion of materials is required, and this is what is tabulated below. Specifically, we need information on

contraction from ambient (conventionally 293K) to the operating temperature of 60K, and to 80K, which is the temperature at which most mechanism testing will be done. Also, values of the CTE at 60K are given for analysis of differential temperature effects. 

More complete information on the CTE is given in section 3. Revision status is given in section 4. 

2. CTE Tabulation


 

Material

 

Contraction, 293-80K (%)

Ave. CTE,

293-80K (ppm/K)

Contraction, 293-60K (%)

Ave. CTE

293-60K (ppm/K)

CTE at 60K (ppm/K)

6061 T651 Al

0.393

18.45

0.408

17.51

5.58

 

A356 Al

0.366

17.18

0.378

16.22

5.2

 

OFHC Cu

0.301

14.13

0.315

13.52

5.05

 

Naval Brass (46400)

0.382

17.92

0.402

17.25

5.0

 

G-10CR (glass fiber/epoxy), parallel to fibers

0.211

9.91

0.223

9.57

5.2

 

G-10CR (glass fiber/epoxy),

perp. to fibers

0.638

29.95

0.667

28.63

12.4

 

Vespel SP3

0.725

34.0

0.761:

32.7:

15::

 

302 stainless steel

0.287

13.47

0.297

12.75

3.8:

 

303 stainless steel

0.282

13.24

0.293

12.58

3.8:

 

440C stainless steel

0.177

8.31

0.182

7.81

1.8:

 

1075 carbon steel

0.187

8.78

0.192

8.24

2.2:

 

BaF2

 

0.313

14.69

0.325

13.95

4.8

CaF2

 

0.296

13.90

0.304

13.05

3.1

MgF2 (par. c-axis)

0.189

8.87

0.192:

8.2:

1.0:

 

MgF2 (perp c-axis)

0.103

4.84

0.104:

4.5:

0.5:

 

Schott SF6 glass

0.1482

6.96

0.1583

6.79

4.86

 

Schott SF57 glass

0.152:

7.1:

0.162:

6.95:

5.0:

 

Schott BK7 glass

0.1203

5.65

.1272

5.46

3.2

 

3. Sources and Discussion

3.1 Aluminum

3.1.1 6061 T651

The data used for the table are taken from the IR/EO Handbook, 3, 358 (1993). These values are ultimately derived from a single measurement by Arp et al, Cryogenics, 2, 230 (1962). Data in other handbooks appear to trace back to this same reference. The only other measurement that I could find is in the Cryogenics Materials Data Handbook (1970), which shows an average CTE about 4% less than the above values. The Arp et al sample is a rod, whereas the second sample is a plate measured longitudinally. The values given in MIL-HDBK-5G are a little more than 1% lower (average of 17.28 ppm/K to 60K); the sources of the data are not given. 

Comparison with measured values for other alloys suggests that a somewhat smaller value may be more representative, but in the absence of more conclusive data I have chosen to follow general practice. It is likely that the observed variation is due in large part to real variations between samples and not just to measurement errors. For this reason, the GNIRS design should avoid situations in which knowledge of the CTE to better than about 2% is critical. 

Some relevant data are plotted below. (The curve from the Cryogenics Materials Data Handbook is not plotted because it is not available in tabular form.) 

3.1.2 A356

Data for A356, which is an alloy commonly used for casting, are also shown. This was because we considered the consequences of casting the GNIRS cryogenic structure out of this material. A356 has a silicon content of 7%, and has a CTE which is lower than that of 6061 by roughly the same amount. 

The cryogenic data on A356 are less extensive. The data plotted and tabulated are from Arp et al; the MIL-HDBK-5G curve is almost 2% lower for the contraction to 60K. 

3.2 Copper and Brass

3.2.1 OFHC Cu

The CTE for OFHC copper is known fairly accurately, certainly to better than 1%. This is probably because it is a nearly pure material. A curve is plotted below. The OFHC data used are taken, again, from the IR/EO Handbook, 3, 359. The pure copper curve is from Materials at Low Temperature, ed. R. P. Reed & A. F. Clark (1983), but is not significantly different from the pure copper curve in the IR/EO Handbook


3.2.2 Naval Brass

This is also known as 46400. This is used in GNIRS mainly for gears. It has a composition that is roughly 60% Cu, 40% Zn and 0.7% Sn. There were no data for material of this exact composition down to 60K. Down to 77K there care CTE data for simlar materials, and these were averaged and interpolated to get the 80K value. Data from Arp et al for 70:30 brass were used to extrapolate to 60K. The CTE ratio of 60:40 to 70:30 brass was taken as 1.13 for this purpose. The overall uncertainty in the CTE values (judging from the scatter in the data at 77K) is about 4%. The higher CTE of brass appears to be due to the Zn, which has a much higher CTE than that of Cu. 

3.3 G-10CR

G-10 is a composite material of glass fibers in a polyester resin. This means, first of all, that it is not isotropic, but in addition one would not expect its properties to be particularly uniform. G10-CR is a cryogenic grade of G-10, which is somewhat more tightly specified (see Kasen et al., Adv Cryo Eng, 26, 235 [1980]) 

The data in the table are taken from Materials at Low Temperature, ed. R. P. Reed & A. F. Clark (1983). These data are also presented by Kasen et al. R. B. Scott (Cryogenic Engineering, 1959) gives values for “molded ployester rod reinforced with glass fiber” that is about 20% larger than the CTE parallel to fibers quoted above. Additional curves can be found in the Cryogenic Materials Data Handbook and elsewhere, including manufacturer’s literature for room temperature. 

The data are plotted below. 

3.4 Vespel

The contraction to 80K is based on a measurement by Steve Rath of contraction between room temperature and liquid nitrogen. The measurement is dominated, undoubtedly, by systematic errors. 

The extrapolation to 60K is based on noting that plastics in the literature typically contract additionally, in going to 60K, by about 5% of the contraction down to 80K. 

The CTE at 60K is a very rough estimate, and is simply a rough average of composite/plastics with similar overall contraction. It is probably good to 5 ppm/K. 

3.5 Steel

It is important to note that the CTE values of steels vary significantly from one alloy to another – more than for aluminum – so it is important to check which alloy is being used. Values are tabulated for 302 and 303 stainless, which are very similar, and for 44C stainless and for 1075 carbon steel.

3.5.1 302 Stainless

The data given in previous versions of the SDN were taken from the IR/EO Handbook. However, although this refers back to the Touloukian et al. volumes, the handbook tabulation does not seem to correspond to any of the values given there. The values now listed are taken from Arp et al, and are corroborated by data from the Cryogenic Materials Handbook. The CTE values are now about 7% larger. 

3.5.2 303 Stainless

The CTE values for this alloy are similar to those for 302 stainless. Note that other 300-series alloys have similar CTE values, although the values may differ from one another by several percent. Data are taken from Arp et al., and corroborated by the Cryogenic Materials Handbook

3.5.3 440C Stainless

This is used in most of the GNIRS bearings. The CTE of this (and other 400 series alloys) is lower than that of the 300 series alloys. 

3.5.4 1075 Carbon Steel

These are data from Arp et al. Carbon content is ~0.8%. Carbon steel (as opposed to alloy steel, which contains substantial amounts of other elements) has a CTE that it close to that of pure Fe; the CTE of the latter is about 4% higher than the values tabulated for 1075 steel, and this suggests that the CTE for a lower-carbon steel will be slightly higher than the values in the table. 

3.5 Optical Materials

Data are given for those IR optical materials used in GNIRS. These comprise several fluoride materials and some Schott glasses. 

3.6.1 Barium Fluoride

The primary source used for these values was NBS Technical Note 993 (Feldman et al, 1978), supplemented by data quoted in Touloukian et al Thermophysical Properties of Materials. Note that Touloukian et al give a “recommended” curve which was not used, although it is plotted below; this is different from the tabulated data that were used. 

Note that these values differ slightly from those used in the initial optical analysis. The average CTE is about 0.65 ppm/K larger than the value previously used, which affects radii and thickness at a level of 0.015%. I am note sure what the difference is due to, probably a combination of the additional data and the extrapolation to 60K. 

The average CTE values are probably good to about 1%, but the value at 60K is probably not much better than 1 ppm/K. 


 
 
 
 

3.6.2 Calcium Fluoride

There are more data available for calcium fluoride. These include the source cited for BaF2 plus Browder and Ballard (Applied Optics, 8, 793 [1969]). The latter source gives measurements for IRTRAN materials, which are not single-crystal materials, and may thus have slightly different properties. The data are nonetheless fairly similar. 

The values given above are, again, somewhat higher than those used in the initial optical analysis, this time by about 1.15 ppm, which will affect dimensions at the 0.03% level. 


 
 

Again, the uncertainties are about 1% in the average CTE values and perhaps 0.5 ppm/K in the CTE at 60K. 

3.6.3 Magnesium Fluoride

MgF2 is an anisotropic material, with a CTE that depends on directions. Both values are given and plotted below; they come from Feldman et al (1978) only since the equivalent IRTRAN material measured by Browder and Ballard is a mixture of material at all orientations. The Ballard and Browder data are used to help extrapolate to 60K, since the CTE there is quite uncertain (see curves below). 

This material is used only for the GNIRS Wollaston prism, so precise knowledge of the CTE is less critical. However, it should be noted that the difference in the contraction along the 2 axes is almost 0.1% to 60K, which corresponds to several thousandths over the width of the prism elements. 


 
 

3.6.4 Schott SF6 Glass

These data were measured for NOAO and CFHT by Schott (Germany) between –180 and 100C.

3.6.5 Schott SF57 Glass

No cryogenic data were available for this glass. The CTE near room temperature is roughly 2.5% greater than for SF6, so values were adopted that were 2.5% greater overall. This assumption is adequate given that the SF57 glass is used only for cross-dispersion prisms. A 10% uncertainty in the CTE, which is plausible, has only minimal effects on the overall design.

3.6.6 Schott BK7 Glass

Again, no cryogenic data were available for this glass (somewhat surprisingly). However, Molby (JOSA 39, 600 [1949]) measured several glasses to ~77K, including two with similar properties (index, dispersion) whose room temperature CTE values bracket those of BK7. An average of the two glasses (BSC-1 & BSC-2) was used that was weighted to have the same room temperature CTE as BK7. Some other curves were plotted in order to see whether a match to SF57 (see above) could be obtained, but none of the glasses measured by Molby were as close as SF6.

 4. Revisions

Revision A, 10/07/99 – Added to discussion of G-10. Correction to discussion of change from previous values of CaF2. No changes to tabulated CTE values.

Revision B, 10/29/99 – Added A356 data. Added discussion of MIL-HDBK-5G data. Corrected some typographical errors. No changes to tabulated CTE values.

Revision C, 12/27/99 – Added brass and steel data. 302 stainless data have changed; see discussion.

 



 

If you have any questions or suggestions regarding this website, please contact Melissa Bowersock.


 

NOAO Intranet Services

NOAO Copyright
 Statement

National Optical Astronomy Observatories, 950 North Cherry Avenue, P.O. Box 26732, Tucson, Arizona 85726,
Phone: (520) 318-8000, Fax: (520) 318-8360